Heat metering refers to the measurement and charging of heat based on heat metering instruments, with central heating or regional heating as the premise, through certain charging policies or heat distribution methods. In layman’s terms, it means paying heating fees based on the heat value measured by the heat metering instrument.
High precision measurement, strong anti-interference ability, good stability.
Protection level reaches IP68, can work in water-immersed and humid environment for a long time (split protection level IP67).
No mechanical rotating parts, no wear, no pressure loss, long service life.
Support RS485 interface, MODBUS communication protocol, support 4-20mA, pulse output and other signal outputs.
Performance | Parameters | |
Nominal diameter | DN15~DN1000 | |
Pipe material | Carbon steel, stainless steel | |
Temperature range (℃) | 4~160 | |
Temperature difference range (K) | 3~70 (factory setting minimum temperature difference 0.2) | |
Minimum pairing temperature error (℃) | ±0.1 | |
Maximum allowable working pressure (Mpa) | 1.6MPa (Special requirements are required if the range exceeds this range) | |
Genauigkeit | Grade 2 | |
Temperature sensor type | PT100, PT100 with pressure installation | |
Schutzniveau | Host | Wall-mounted host: IP67 Integrated host: IP68 |
Flow sensor | IP68 | |
Temperature sensor | IP68 | |
Working power supply | DC8~36V, AC85~264V | |
Power consumption | 1.5W | |
Working environment | Temperature: -20~75℃℃; Humidity≤100%(RH) | |
Communication and output | RS485, 4-20mA, cumulative pulse, wireless | |
Local display | Dual-line display of cumulative amount, instantaneous flow and instrument working status | |
Display resolution | Accumulated heat 0.001KW.h; thermal power: 0.0001KW; temperature: 0.001℃; temperature difference: 0.0001K | |
Maximum reading | Net cumulative flow -9999999~9999999m3 | |
Net accumulated heat 0~9999999KW.h | ||
Storage temperature (℃) | -20~+60 | |
Data storage | Wall-mounted host: SD card timed storage of set parameters and measurement results (optional) | |
Instrument installation location | Water supply pipeline, return pipeline |
1. Split Host
Wall-mounted ⅠHost Installation Instructions
The wall-mounted type is used for wall installation and is fixed with 4 φ6mm expansion bolts or nails.
The rail fixing clip is used for rail installation.
Use PCB mounting bracket for rail installation
The wall-mounted host can be installed on the wall, distribution box and explosion-proof box using rail installation.